High frequency radar level meter Installation Requirements - Just Measure it

High frequency radar level meter Installation Requirements

Installation guide:

Be installed in the diameter of 1/4 or 1/6.

Note: The minimum distance from the tank wall should be 200mm.

Note: ① datum

②The container center or axis of symmetry.

The top conical tank level can be installed at the top of the tank is intermediate, can guarantee the measurement to the conical bottom.

A feed antenna to the vertical alignment surface.

If the surface is rough, the stack angle must be used to adjust the angle of Cardan flange of the antenna to the alignment surface. (Due to the solid surface tilt will cause the echo attenuation, even Loss of signal.)

Typical installation errors

The conical tank cannot be installed above the feed port.

Note: outdoor installation should adopt sunshade.

1 Correct

2 Error rainproof measures.

The instrument cannot be installed in the arched or domed roof intermediate. In

addition, to produce indirect echo is also affected by the echoes. Multiple echoes can be larger than the real value of signal echo because through the top can

concentrate multiple echoes. So cannot be installed in a central location.

There are obstacles affecting measurement needed reflection plate.

Height of nozzle:

The antenna extends into the tank at least a 10mm distance.