Connect two-wire or three-wire, three-wire from the terminal AGND (GND) lead break wire with the case short or the black jumper cap of the rear panel from ON to OFF end, or remove the signal line at the end of the instrument, a single measurement of the instrument end of the current output is normal

Gas turbine flow meter fault analysis – meter shows 0 with no flow but current output Read More »

Turbine flowmeter fault – flow indication when not running

 ● Whether it is an industrial frequency disturbance (50HZ), characteristic: a value is displayed that does not change; this value can be calculated according to the following formula q—instantaneous flow rate (m3/h) f — frequency (HZ) K — coefficient (times/m3) ● Whether the wiring is laid in a line slot with strong electricity (220V or

Turbine flowmeter fault – flow indication when not running Read More »