Flow meter

Gas turbine flow meter failure analysis-there is display when no flow

● Interference problem: Ask the instrument if there is an external power supply, if there is, remove both the external...
Read More
Flow meter

Gas turbine flow meter fault analysis – pressure display 0 or inaccurate

The pressure range selection is large, and the small field pressure cannot be collected. The pressure calibration parameters in the...
Read More
Flow meter

Gas turbine flow meter fault analysis – Abnormal temperature display

If the temperature display is abnormally high when the battery is powered, the battery is dead. If the temperature display...
Read More
Flow meter

Gas turbine flow meter fault analysis – no pulse output

● Since D2 is a dual power supply, to determine whether 24V is properly powered, unplug the battery or test...
Read More
Flow meter

Gas turbine flow meter fault analysis – no pulse equivalent output

● Since D2 is a dual power supply, to determine whether 24V is properly supplied, unplug the battery or test...
Read More
Flow meter

Gas turbine flow meter fault analysis – meter shows 0 with no flow but current output

Connect two-wire or three-wire, three-wire from the terminal AGND (GND) lead break wire with the case short or the black...
Read More
Flow meter

Gas turbine flow meter fault analysis – screen does not light up

Measure if the power supply voltage is normal or if the battery (not less than 3V) is charged at the...
Read More
Flow meter

Turbine flowmeter fault – meter slowly zeroing in operation

● Worn or damaged movement● The flow rate becomes smaller due to a gradual lack of pressure in the pipeline...
Read More
Flow meter

Turbine flowmeter fault – flow indication when not running

 ● Whether it is an industrial frequency disturbance (50HZ), characteristic: a value is displayed that does not change;this value can...
Read More
Flow meter

Turbine flowmeter failure – 485 communication not working

Default address 001, baud rate 9600, n without parity, 8 bits of data, 1 bit of stop Is the instrument...
Read More
Flow meter

Turbine flow meter fault – no pulse output

● Pulse mode and pulse equivalent can be modified in the parameters, refer to the manual for details ● Pulse...
Read More
Flow meter

Turbine flowmeter fault – current zero point drift

● The blue potentiometer can be adjusted by unscrewing the front cover directly; see pictureLeft zero, right full4mA potentiometer on...
Read More
Flow meter

Turbine flowmeter fault – screen does not light up

● Measure if the voltage supply to the instrument is normal or if the battery is charged ● Consider whether...
Read More
Flow meter

Turbine flowmeter fault – current output abnormal

(1) Three-wire current, current output is not normal: ● Pay attention to the wiring: 24V, COM (GND), IOUT ● Measure...
Read More
Flow meter

Gas turbine flow meter fault-485 communication problems or no communication

The default address of the instrument is 001, baud rate 9600, n without checksum, 8 bits of data, 1 bit...
Read More
Flow meter

Gas turbine flow meter fault-fast flow display

Whether the customer has moved the parameters Is the installation in accordance with the requirements Whether the flow rate is...
Read More
Flow meter

Gas turbine flow meter fault-Slow flow display

● Whether the customer has set the parameters ● Is the installation in accordance with the requirements ● Whether the...
Read More
Flow meter

Turbine flow meter fault-Abnormal temperature/pressure display

Abnormal instrument temperature displayIf the temperature display is abnormally high when the battery is powered, the battery is dead.If the...
Read More
Flow meter

Vortex flow meter fault -display 0 when meter has flow

Ask if the flow rate at the site is lower than the lower limit, then change the small flow rate...
Read More
Flow meter

Vortex flow meter fault-Incorrect flow rate display

● Check that the temperature and pressure of the instrument are displayed properly ● Ask the customer about the inaccuracy...
Read More