Metal float flowmeter common troubleshooting methods - Just Measure it

Metal float flowmeter common troubleshooting methods

 

Phenomenon

Reason

Treatment Measures

Float flowmeter pointer jitter 1. Medium pulsation. 1. Increase the damping.
2. Unstable air pressure. 2. Using voltage or current stabilization device.
The pointer stops at a certain position and does not move 1. Open the valve too fast and the float is stuck. 1. Slow down the opening speed of the valve.
2. The float guide rod is not centered with the stop ring. 2. Remove the meter and adjust it.
Large measurement error 1. The installation does not
meet the requirements.
1. Vertical or horizontal installation inclination
is not more than 20 degrees.
2. The density of the liquid
medium varies widely.
2. Calculate the error correction factor and multiply the
reading to convert it to the true flow rate.
3. Temperature pressure has
a large impact.
3. Adopt temperature and
pressure compensation.
4. Pipeline vibration. 4. Find a professional to adjust the
position of the components.