Turbine flow meter fault - no pulse output - Just Measure it

Turbine flow meter fault – no pulse output

● Pulse mode and pulse equivalent can be modified in the parameters, refer to the manual for details

● Pulse active output, PNP, NPN, and OC hardware can be commissioned at the rear terminal board, refer to the manual for details

● Use a multimeter to measure the voltage between FOUT-C and COM, under the condition that the instrument has flow, if it measures 0V (low level) or about 24V (high level), it means that the instrument has no output and the circuit board is faulty;

 if it measures a voltage between 10V-12V, it means that there is an output (supply 24V, half of the voltage is measured by the multimeter when there is an output), the customer can check the acquisition equipment there is no problem.