Target flowmeter Index and Display - Just Measure it

Target flowmeter Index and Display

Main index and display

  1. Index
Medium Liquid, Gas ,Steam, and LPG
Caliber Pipe type 15~300mm Clamp type 15~600mm Plug type100~2000mm
Pressure 0.6~10MPa 0.6~42MPa 0.6~42MPa
Medium Temperature -196℃~+450℃ ( Confirming the temperature range before purchase)
Precision ±0.2% ±0.5% ±1.0% ±1.5% ±2.5%
Range 1:03 1:05 1:10 1:10 1:10Steam
Compensation Temperature compensation; Pressure compensation
Repeatability 0.1%~0.08%
Power Supply 3.6V lithium battery, 24V DC
Output LCD display, and 4~20mA, 0~10V, impulse or RS485 (Optional)
Material of measure tube Carbon Steel;304;316L; Being able to provide others according to user requests
Explosion- Prevent Symbol ExiallCT4, ExiallBT4
Protection Rank IP65;IP67

2. LCD display

A. Run in normal

(Temperature or pressure been not displayed respectively if no setup temperature compensation or pressure compensation.)

B. Run in debug

LCD displays the following contents in debug process:

Adjustable Coefficient

Flow Coefficient:CL××××

Pressure Coefficient:CP××××

Temperature Coefficient:Ct××××

Serial No.:n××

20mA or 10V outputs corresponding to the maximum flow: H×××××

Zero-temperature compensation coefficient:LT×××

Nonadjustable Coefficient:

Zero- coefficient of flow sensor:EL×××××

Zero- coefficient of pressure sensor:EP×××××

ZeroInstrument

Share This Story, Choose Your Platform!

Contact Us

    Please prove you are human by selecting the car.