1) The flow meter display non-zero, while the flow velocity is zero. The following reasons and solution methods may be included:
- Flowmeter was not installed in a strictly horizontal position. Solution: zero set operation.
- Flow meter has run for a long time and error occurs. Solution: zero set operation.
- Overloading brings zero offsets. Solution: zero set operation.
- The sensor is wetting or no insulation. Solution: drying sensor by drier or return the flow meter to the factory.
- Some eye-winker clips the target or transition part. Solution: close front and behind valves and
- loosen the connecting bolt between transitional part and measure tube, then sway lightly the
- target or transition part or put them out and clean them, finally replace them.
- A high-temperature type flow meter without temperature compensation has not been set zero under
- high-temperature conditions. Solution: zero set operation under high-temperature conditions.
2) LCD display value appears non-normal accretion. The following reasons and solution methods may be included:
- Some silk or belt-type eye-winker accretes on the target. Solution: clean the target.
- Much dirt accretes on the target. Solution: clean the target.
- The sensor is wetting then the insulation resistance is decreased. Solution: drying sensor by drier or return the flow meter to the factory.
3) Measurement error big. The following reasons and solution methods may be included:
- The flow meter is installed in a non-coaxial position corresponding to the connecting pipeline. Solution: adjust the installation position.
- The front or behind connecting pipe is so short. Solution: re-installation or re-test.
- The by-pass pipe is of leakage. Solution: check or replace it.
- Some silk or belt-type eye winker accretes on the target. Solution: clean the target.
- Integration microcomputer circuit board runs wrong. Solution: Return the flow meter to the factory.
4) LCD displays nothing or invariable numbers. The following reasons and solution methods may be included:
- Battery or 24V DC power is connected wrong or bad. Solution: check and re-connect all of the connecting wires and clips.
- The Circuit board or LCD is damaged. Solution: return the flow meter to the factory.
5) LCD displays zero always. The following reasons and solution methods may be included:
- The target falls off or is damaged. Solution: Re-install the same type of target.
- The sensor is damaged. Solution: Return the flow meter to the factory.
- Some eye winker clips between target and transition part. Solution: Clean them or return the flow meter to the factory.
- Flow velocity is so low to lees than the minimum flow scale. Solution: Return the flow meter to the factory to replace another type of target.
6) Output does not run or runs wrong. The following reasons and solution methods may be included:
- Output connection is bad or wrong. Solution: Check and re-connect all of the connecting wires and clips.
- The Circuit board is damaged. Solution: Return the flow meter to the factory.
- The user’s signal receiving system runs bad or wrong. Solution: Check and debug it.
General Notice
- LCD glittering means the power of the battery is going to exhaust, please replace the same type of battery.
- Flowmeter stores automatically all of the measure values and sets coefficients in its inner FLASH memory.
- If the environmental temperature is below -20℃, the LCD will be frozen and display nothing. But it will return to normal work at once when the temperature returns up -18℃.