Magnetic flow meter troubleshooting - Just Measure it

Magnetic flow meter troubleshooting

Here list of some most common faults of the magnetic flowmeter and the troubleshooting.

1.  No display No power Apply correct power
Fuse blown Replace a fuse with the same parameter
The contrast of LCD is too low Increase the contrast
2. Empty Pipe alarm Fluid is not full filled the pipe Increase the flow rate
Electrode was polluted Clean the electrode if the voltage of DS1 and DS2>1V
The fluid’s conductivity is too small If connect three terminals SIG 1, SIG 2, SGND and the alarm disappears,
which means the fluid’s conductivity is small. Replace another kind of flowmeter
3. Flow rate indication is unstable Grounding issue Make sure the meter is properly grounded to
a good earth ground
Air Make sure the fluid does not contain air bubbles
Converter location outside
electrical interference
Make sure the converter is not too close to sources
of electrical interference

 

Share This Story, Choose Your Platform!

Contact Us

    Please prove you are human by selecting the cup.