Magnetic flow meter straight length requirements - Just Measure it

Magnetic flow meter straight length requirements

Required lengths of straight pipe

For optimum accuracy performance, it is required to provide sufficient inlet and outlet straight pipe runs. An equivalent to 3 diameters of straight pipe is required on the inlet side, and 2 diameters on the outlet side.

There are no special requirements for standard concentric pipe reducers. See diagram 1 for required straight runs when there is an altering device.

 

When the meter contains removable cover plates, leave the cover plate installed unless accessory modules specify removal.

Don’t remove the cover plates when the meter is powered, or electrical shock and explosion hazard can be caused.

Share This Story, Choose Your Platform!

Contact Us

    Please prove you are human by selecting the cup.