Electromagnetic flow meter failure-no pulse output - Just Measure it

Electromagnetic flow meter failure-no pulse output

Fault point: Converter

Test tool.

Verification scheme.

1. Confirm if there is a pull-up resistor at the site

2. Whether there is the voltage at the pulse acquisition end

3. Check if the wiring is correct

Test result: 1.

1. No pull-up resistor

2. No voltage

3. Wrong wiring

Treatment solution: 1.

1. Pull up a 3K resistor

2. Check if the wiring is correct

3. Correct wiring (if there is a problem, please consult our  after-sales staff)

Share This Story, Choose Your Platform!

Contact Us

    Please prove you are human by selecting the heart.