Turbine flow meter fault-Abnormal temperature/pressure display

Abnormal instrument temperature display If the temperature display is abnormally high when the battery is powered, the battery is dead. If the temperature display is below zero or much lower, the circuit board temperature chip is damaged, return the circuit board to the factory or send a replacement board. Use a multimeter to measure the

Turbine flow meter fault-Abnormal temperature/pressure display Read More »

Vortex flow meter fault- 485 communication cann’t connected

485 communication cannot be connected The parameter confirmation function is selected for the 485 functions refers to the manual The default address of the instrument is 001, baud rate 9600, n without parity, 8 bits of data, 1 bit of stop Is the instrument wired correctly, connect 24V COM 485A 485B, battery power supply and

Vortex flow meter fault- 485 communication cann’t connected Read More »