Small Signal Cutoff in Process Instrumentation: A Practical Guide - Just Measure it

Small Signal Cutoff in Process Instrumentation: A Practical Guide

What Is Small Signal Cutoff?

Have you ever encountered these situations?

  • The pipeline is completely stopped, but the flowmeter still displays 0.2 m³/h.
  • The process has been shut down, yet the totalized flow continues to increase.
  • The DCS constantly shows a tiny flow value even though there is no actual flow.

In many cases, the instrument is not faulty. The real cause is that Small Signal Cutoff (Low Cut) has not been configured correctly.

Small Signal Cutoff is a software function that filters false measurements caused by electrical noise, zero drift, or signal fluctuations near zero. It is primarily used for flow measurement, especially differential pressure flowmeters (orifice plates, wedge flowmeters, and V-cone flowmeters). It is generally not recommended for pressure, level, or temperature measurements.

Most control systems only provide Low Cut (Lower Small Signal Cutoff). A few advanced systems also support High Cut (Upper Small Signal Cutoff), although it is rarely used in industrial applications.

1. Low Cut (Lower Small Signal Cutoff)

Definition

If the measured value is below the configured Low Cut threshold, the system forces the process variable (PV) to zero.

Logic

  • Measured Value < Low Cut → PV = 0
  • Measured Value ≥ Low Cut → Normal measurement and totalization

Why Is Low Cut Necessary?

Low Cut helps eliminate false measurements caused by:

  • Zero drift when there is no actual flow
  • Electrical interference and signal noise
  • Amplified errors after square-root extraction in differential pressure flowmeters
  • False totalization during equipment shutdown

Without Low Cut, a flowmeter may continue accumulating non-existent flow over long periods, resulting in significant measurement errors.

Typical Low Cut Settings

Flowmeter TypeRecommended Setting
Orifice Plate / Differential Pressure3–8% FS
Wedge Flowmeter3–8% FS
V-Cone Flowmeter3–8% FS
Electromagnetic Flowmeter1–5% FS
Vortex Flowmeter1–5% FS
Ultrasonic Flowmeter1–5% FS
Custody Transfer ApplicationsKeep as low as practical

Note: For custody transfer applications, Low Cut should be configured carefully to avoid under-measurement of legitimate low flow rates.

Example

Flow Range: 0–100 m³/h

Low Cut: 3 m³/h

  • Flow = 2.5 m³/h → Display = 0
  • Flow = 3.2 m³/h → Display = 3.2 m³/h

2. High Cut (Upper Small Signal Cutoff)

Definition

High Cut is an optional function available in a small number of advanced control systems.

When the measured value exceeds the configured threshold, the system may:

  • Clamp the value at the configured upper limit, or
  • Ignore measurements above the threshold.

Important Notes

High Cut is not the same as a High-High Alarm.

Unlike alarms, High Cut directly affects the measured process value.

Typical Applications

High Cut is only suitable for a few specialized situations, such as:

  • Special protection loops
  • Micro-flow monitoring systems
  • Laboratory or research applications

For over 95% of industrial flow measurement applications, High Cut is unnecessary.

Do not enable High Cut for normal process flow measurement, as it may suppress valid measurement data.

3. Commonly Confused Concepts

Low Cut ≠ Lower Range Value

Lower Range Value (LRV)

Defines the starting point of the instrument’s measurement range.

Low Cut

Only creates a software threshold for filtering measurements and does not change the instrument’s actual measuring capability.

Low Cut ≠ Alarm Limits

Low Cut

Changes the displayed and calculated process variable (PV).

High/Low Alarm

Only generates an alarm notification while leaving the measured value unchanged.

Differential Pressure Flowmeters: Correct Processing Sequence

The correct processing order is:

Low Cut → Square Root Extraction

Never reverse the sequence.

Square-root extraction amplifies very small differential pressure signals. If square-root extraction is performed before Low Cut filtering, tiny noise signals near zero may appear as measurable flow, leading to inaccurate totalization.

4. Practical Considerations

Custody Transfer Measurement

Do not configure an excessively large Low Cut value.

Any adjustment should be reviewed and approved by process personnel to avoid under-measurement.

Safety Instrumented Systems

Use Low Cut with caution for measurements participating in safety interlocks.

If a low-flow protection function relies on the measured PV, forcing the PV to zero may result in unintended trips.

Transmitter vs. DCS Configuration

Enable Small Signal Cutoff in only one location, preferably in the DCS.

Enabling it simultaneously in both the transmitter and the DCS may result in:

  • Double filtering
  • Unexpected value jumps
  • Difficult troubleshooting

Pressure, Level and Temperature Measurements

Small Signal Cutoff is generally not recommended.

Minor fluctuations in these measurements often represent actual process conditions.

Filtering them may hide slow leaks, gradual level changes, or other important process information.

5. Common Troubleshooting Cases

Problem 1

A small flow exists, but the DCS always displays zero.

Cause

The Low Cut threshold is configured too high.

Problem 2

The equipment is stopped, but the totalized flow continues increasing.

Cause

Low Cut is disabled, allowing zero drift and signal noise to accumulate.

Problem 3

The displayed value repeatedly switches between zero and the actual flow.

Cause

No hysteresis has been configured.

A typical solution is:

  • Low Cut = 3%
  • Recovery Threshold = 3.5%

This prevents frequent switching when the measured value fluctuates around the cutoff point.

Key Takeaway

Small Signal Cutoff does not improve instrument accuracy.

Its purpose is to prevent false signals near zero from participating in process indication and totalization.

A Low Cut setting that is too small allows false flow accumulation during shutdown.

A setting that is too large may suppress legitimate low flow measurements.

Therefore, the optimal Low Cut value should always be determined according to the specific process conditions and application requirements—not simply set as high as possible.

Share This Story, Choose Your Platform!

Contact Us

    Please prove you are human by selecting the star.
    Translate »